Conclusion a procedure for monitoring of cmms with

Info iconThis preview shows page 1. Sign up to view the full content.

View Full Document Right Arrow Icon
This is the end of the preview. Sign up to access the rest of the document.

Unformatted text preview: to GUM the measurement uncertainty is calculated with the VCMM3 software, which bases on Monte-Carlo simulations. The shaft collet was calibrated at the reference CMM Zeiss UPMC 1200 CARAT S-ACC in several positions to get traceable results (figure 2). The measurement results are documented in the calibration certificate together with the feature specific extended measurement uncertainties. Figure 2; DKD calibration of a shaft collet at the reference CMM 4.2. Segmenting the CMM measuring range Even to allow for a statement about the capability of the measurement process of a large CMM like a horizontal-arm CMM, its measuring volume is divided into twelve monitoring fields in (figure 3). The shaft collet was measured at six positions in the x-y-plane and at two z-levels. 1 2 Physikalisch-Technische Bundesanstalt (PTB), the German national metrology institute Deutscher Kalibrierdienst (DKD), association of calibration laboratories 3 Virtual Coordinate Measuring Machine (VCMM) 132 z y y x x A. Weckenmann, S. Beetz and J. Lorz monitor fields monitoring fields measurement plate ma...