View Full Document Right Arrow Icon
Analysis of Ca in Foods with PerkinElmer AAnalyst100 AA Spectrometer smk 11.2008 DO NOT PULL the fine aspirator tube. Supply deionized water to the tube when the instrument is idle. After every sample, wipe tube with Kimwipe tissue to avoid cross-contamination. Complete the information requested on the “AA Printer Sheet” (downloaded from WebCT › Retrieve Information) and feed it to the printer. Turn on the printer. On the printer display panel, the (select) light must be on; if not, press the [ Select ] button. Press [ Print FF ] on the spectrophotometer keypad to engage printer. 1. Printer check: While supplying deionized water to instrument, press the white [ Data ] button and the [Read] button on the instrument keypad. If printer does not function, ask for assistance. 2. Zero the instrument: ¾ Press the white [ Cont ] button to select continuous reading mode. ¾ Supply your blank solution to the instrument & press the [ A/Z ] (auto-zero) button. ¾