2 The horizontal error bars represent the 01 mA uncertainty in the True Current

2 the horizontal error bars represent the 01 ma

This preview shows page 6 - 10 out of 10 pages.

Click on the black triangle by the Curve Fit icon on the graph toolbar and select Linear. 2.The horizontal error bars represent the +/- 0.1 mA uncertainty in the True Current which was achieved 3.The uncertainties in the slope and intercept arise from the spread of the data points but do not includethe uncertainty in the True Current. 4.Answer the first two Questions on the Conclusions page.Conclusions:Ohm’s Law1.How well does your data support Ohm’s Law? Explain fully!2.What is the physical meaning of the slope of the Linear Fit to the data on the Ohm’s Law graph. Hint: what are the units of the slope?
Image of page 6
Page 7 of 9 Equivalent Circuits Diagrams: Figure 1: Series Figure 2: Parallel Figure 3: Series/Parallel Figure 5: Series Figure 6: Parallel Figure 7: Series/Parallel
Image of page 7
Page 8 of 9 Equivalent Resistance: Resistor Check: the resistors on the UI-5210 circuit board are accurate to within +/- 5%. This can be improved substantially if a multi-meter is available. These will generally measure resistance +/- 1%. 1. Disconnect all the wires from the circuit board. 2. Use a multi-meter to measure the resistance of resistors R2, R3, & R4 on the circuit board. Enter the values in the Resistor Check Values table, replacing the nominal values in column 2. Theory Resistance: Using Equivalent Circuits (see Theory section above) and the values for the resistors from the Resistance Check Values table, calculate the equivalent resistance for each of the three circuits shown on the previous page. Enter the values in the Theory Resist. column of the Equivalent Circuits table below. Resistor Resistance (Ω) Circuit Theory Resist. (kΩ 1 R2 3300 1 Series 2 R3 1000 2 Parallel 3 R4 1000 3 Series/Parallel 4 4 Resistor Check Values Table Equivalent Resistance Table
Image of page 8
Page 9 of 9 Procedure: Equivalent Circuits 1. Click open the Signal Generator. 2. Under Output 1: a. Set waveform to DC b. Set DC voltage to 15 V c. Set Voltage Limit to 15 V and the Current Limit to 1.5 A d. Press the ON button and close the panel Calibrate Check:
Image of page 9
Image of page 10

  • Left Quote Icon

    Student Picture

  • Left Quote Icon

    Student Picture

  • Left Quote Icon

    Student Picture